1
1010AERWDP TRANSDUCER INSTALLATION
1. INTRODUCTION
The System 1010AERWDP flowmeters are part of Siemens extensive System FUS 1010 Clamp-
On Transit-Time Ultrasonic Flowmeter division. These meters have all the capabilities of the
FUS1010 Clamp-On Series flowmeters plus all the added features of the Aerospace line, which
is specifically designed to accurately and non-intrusively measure flow in steel, aluminum,
and titanium tubing typically used in aircraft hydraulic and fuel systems.
NOTE: If not installing aerospace transducers refer to the FUP1010 Portable Flow-
meter Field Manual instead.
2. FLOWMETER SETUP
NOTE: If flowmeter is already installed at the site disregard the following set up proce-
dure.
The following instructions require the use of the FUP1010 Flowmeter keypad and menus. The
installer should become familiar with them before proceeding further. Refer to Tables 1 and 2
as needed for menu setup data. In addition, refer to Section 2 - The Installation Menu in the
appropriate flowmeter field manual.
To set up the flowmeter with the Aerospace Transducer, proceed as follows:
Selecting Aerospace Transducer
z Switch the flowmeter on. Press the <MENU> key.
z Press <Right> arrow key to select “Dual Channel Flow” mode. To select, press <ENT>
key.
z Press <Right> arrow key and then the <Down> arrow key to select Aerospace, then
press the <ENT> key.
Siemens Site [1]
Dual Channel Flow Setup
Dual Channel Flow >Aerospace
Clamp-on
Flow Tube
Reflexor
Single Channel
Meter Facilities
Siemens Site [1]
Select Meter Type
Meter Type >Dual Channel Flow