6-7
1010PFM-3G
Section 6
To invoke install routine press . This triggers Initial Makeup (an internal process that in-
terrogates the flow tube and analyzes setup data to optimize operation). During procedure,
the meter shows its progress by highlighting the prompt line.
Upon completion, the computer reports the measured sonic velocity. You can accept this read-
ing or press to adjust it (see field manual for instructions).
ENT
You can complete the Installation procedure without actually mounting the flow tube at its intended location. Fill
the flow tube completely with liquid and connect the Up and DN cables according to the Installation/Outline
Drawing. Complete the instructions in this section. If you choose to mount the Flow Tube first, please conform
to the guidelines presented in this section and the Installation Drawing. In any case, do not proceed with these
instructions until the flow tube is filled completely with liquid.
After filling the flow tube with liquid, press to access the [Install Completed?] option list and
then move the cursor to [Install] by pressing
Note that the default Empty Flow Tube Set is [MTYmatic]. The flow computer deter-
mines the empty flow tube setting automatically during the Initial Makeup, so use of
this menu cell is optional.
Siemens 2 Channel [1] FLOW1
Select FlowTube 1S
Install Completed? >Yes
Empty FlowTube Set New Makeup
Zero Flow Adjust Chan Not Setup
Key Install to start operation
Install FlowTube
Siemens 2 Channel [1] FLOW1
Select FlowTube 1S
Install Completed? Install
Empty FlowTube Set Chan Not Setup
Zero Flow Adjust Chan Not Setup
Drive 27 m 10 [6: :0]
Install FlowTube