z Upon completion the meter reports the measured liquid sonic velocity. To accept the
Measured Vs, press the <Down> arrow.
NOTE: If you want to correct the reported sound velocity, press the <Right> arrow to
edit the indicated Vs value, then press <ENT> to continue.
NOTE: The meter will now perform an AutoZero operation where the meter measures
any zero flow offset, which is used to correct the flow reading.
Siemens Site [1] SITE1
Pick/Install Xdcr
Transducer P/N 1011AP-16S2-B
Install Completed? Yes
Empty Pipe Set MTYmatic
Zero Flow Adjust Actual Zero
Set Empty Pipe Detection Threshold
Siemens Site [1] SITE1
Pick/Install Xdcr
Transducer P/N 1011AP-16S2-B
Install Completed? Install
Empty Pipe Set Chan Not Setup
Zero Flow Adjust Chan
Not Setup
Measured Vs m/s
1490
Drive 10 m 13 [6:----------:0]
z Press the <Down> arrow to select Install Completed?, then press the <Right> arrow.
z Use the <UP> or <Down> arrows to move the cursor to Install.
z Press the <ENT> key to register selection. The meter will then perform an Initial Makeup,
which conditions its operation to the selected transducer, pipe, and liquid. It then auto-
matically sets the empty pipe and zero flow levels. Be patient, this process may take
several minutes.
Siemens Site [1] SITE1
Pick/Install Xdcr
Transducer P/N 1011AP-16S2-B
Install Completed? No
*Install
Empty Pipe Set Chan Not Setup
Zero Flow Adjust Chan Not Setup
Key [Install] after mounting transducers