2-16
1010PFM-3GSection 2
2.4 THE PICK/INSTALL XDCR (TRANSDUCER) MENU
Use this menu after creating a new site setup in the Channel Setup menu, and defining the pipe para-
meters in the Pipe Data menu.
Based on pipe data (and optional application data) entries, the Pick/Install Xdcr menu automatically
identifies the most suitable transducers for the application. It recommends the appropriate mounting
mode (direct or reflect) and lists the Spacer Bar or Mounting Track part number and spacing index.
Ideally, you will be able to use the primary recommendations. However, you can edit the menu entries
as required to accommodate different transducers or mounting configurations.
The flow computer will adjust its parameters to optimize performance based on your selections. The
Ltn menu cell shows the required spacing distance (in inches or millimeters) between the upstream
and downstream transducers. Use the [Install Completed?] menu cell to inform the flow computer that
you completed the physical mounting of the transducers. You can define the empty pipe and zero flow
values once the transducers are operational.
NOTE: Before preceeding to mount transducers, it is recommended that Section 3- HARD-
WARE INSTALLATION GUIDE be reviewed. Refer to the Pick/Install Xdcr menu and
menu structure shown below for menu cell descriptions and details.
Siemens 2 Channel [1] FLOW1
Scroll List and select desired Model
Pick/Install Xdcr
Use this menu cell to
select the type of
transducer to be
installed.
This menu cell
allows you to select
a mounting mode.
*Use Reflect Mode
whenever possible.
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Reflect
Spacing Method Spacer Bar PB-1
Letter Index C
Number Index 0
Ltn Value (in) 0.565
Install Completed? No
Empty Pipe Set Chan Not Setup
Zero Flow Adjust Chan Not Setup
These menu cells
identify the spacing
indices on the spacer
bar. You must use
these settings to mount
the transducer properly.
This menu cell indicates the
model number of the
appropriate spacer bar for
your transducers and
spacing requirements.
Once you select the
type, use this menu
cell to specify the
transducer size.
Use this menu cell to
inform the meter that
you completed the
transducer mounting.
A
fter the transducers are
operating, you can use
this menu cell to set the
empty pipe threshold.
This menu cell
shows the actual
spacing distance
required between
the transducers.
After the transducers are operating,
use this menu cell to manually set
the zero flow correction.