Setup
3.4 Inputs and outputs
FST030
136 Function Manual, 04/2017, A5E35693328-AA
Frequency output
3.4.5.6.1 Redundancy mode
Description
Sets the Redundancy mode. The frequency on channel 3 will follow frequency on channel 2
with 90° or 180° shift.
Setting
Off. Channel 3 runs in independently; 90°. Channel 3 follows frequency on channel 2 with a
90° shift; 180°. Channel 3 follows frequency on channel 2 with a 180° shift
Off. Channel 3 runs in independently
Path
Setup → Inputs and outputs → Channel 3 - input/output → Frequency output → Redundancy
3.4.5.6.2 Process value
Description
Select process value for the frequency output.
Setting
Volume flow; Mass flow; Standard volume flow (only hydrocarbon and gas applications);
Flow velocity; Sound velocity; LiquIdent (only hydrocarbon applications); Density; Kinemat-
ic viscosity; Pressure; Medium temperature; Standard density (only hydrocarbon and gas
applications); Standard kinematic viscosity (only hydrocarbon applications); API gravity
(only hydrocarbon applications); Standard API gravity (only hydrocarbon applications);
Specific gravity (only hydrocarbon applications); Standard specific gravity (only hydrocar-
bon and gas applications); Auxiliary temperature; Sensor temperature; Standardizing factor
(only hydrocarbon and gas applications)
Path
Setup → Inputs and outputs → Channel 3 - input/output → Frequency output → Process
3.4.5.6.3 Direction
Description
Flow direction filter applicable for process values mass flow/volume flow/flow
velocity/standard volume flow rate (only hydrocarbon and gas applications).
Positive; Negative; Bidirectional; Bidirectional (symmetric)