Setup
3.6 Local display
FST030
208 Function Manual, 04/2017, A5E35693328-AA
6th process value
Description
The display
can be selected.
Setting
Volume flow; Mass flow; Flow velocity; Sound velocity; Density; Process viscosity; Process
pressure; Fluid temperature; Reference density; Totalizer 1; Totalizer 2; Totalizer 3
This parameter is only available if
is set to Six values or Six diagnostic values.
Setup → Local display → View 1 → 6th process value
Trend scale mode
Description
The
can be selected with automatic scaling or fixed scaling.
This parameter is only available if
is set to 1 value and graph.
If Fixed is selected,
and
must be configured.
Setup → Local display → View 1 → Trend scale mode
Trend log time window
Description
The
logging period (time axis length) can be selected.
1 minute; 5 minutes; 15 minutes; 30 minutes; 1 hour; 2 hours; 3 hours
This parameter is only available if
is set to 1 value and graph.
Setup → Local display → View 1 → Trend log time window