Maintenance and diagnostics
4.2 Diagnostic events
FST030
250 Function Manual, 04/2017, A5E35693328-AA
Related
The selected events will be shown in the display when they occur. Furthermore, they will be
indicated on the output if
Output (2)
,
Input/output (3)
and/or
Input/output (4)
is set to Status
output and
Status mode
is set to Alarm item.
Path Maintenance and diagnostics Diagnostic events → Enable alarms → Input/output events
(1)
4.2.7.12
Input/output events (2)
Description
If one of the selected events appears this affects the status icon, the related status output
and the alarm on the communication interface.
Setting 186 CH3 input current too low; 187 CH3 input current too high; 188 CH3 external failure
Default
Related
The selected events will be shown in the display when they occur. Furthermore, they will be
indicated on the output if
Output (2)
,
Input/output (3)
and/or
Input/output (4)
is set to Status
output and
Status mode
is set to Alarm item.
Path
Maintenance and diagnostics → Diagnostic events → Enable alarms → Input/output events
(2)
4.2.8
Assign alarm class
4.2.8.1
14 - Path 1: No signal
Description
Warning alarms for Path 1
Setting
Process value warning; Maintenance demanded; Maintenance required
Default
Maintenance required
Path
Maintenance and diagnostics → Diagnostic events → Assign alarm class → 14 - Path 1: No
signal
4.2.8.2
15 - Path 2: No signal
Description
Warning alarms for Path 2.