Communication
5.2 HART (Channel 1)
FST030
Function Manual, 04/2017, A5E35693328-AA
311
5.2.4
HART device type (read only)
Description
The
HART device type
identification can be entered. The ID uniquely identifies the device
when combined with the manufacturer identification and device type.
Path
Communication → HART (Channel 1) → HART Device Type
5.2.5
HART device revision
Description
Displays device revision level defined in HART protocol.
Path Communication → HART (Channel 1) → HART revision
5.2.6
Response preambles
Description
The number of response preambles can be defined.
Setting 5 to 20
Default
5
Path
Communication → HART (Channel 1) → Response preambles
5.2.7
Dynamic variable mapping
Mapping of measured process variables
The assignment of the measured process values to HART device variables (PV - primary
variable; SV - secondary variable; TV - tertiary variable; and QV - quaternary variable) can
be modified and assigned as desired via local user interface or via HART interface using
SIMATIC PDM.
PV
The process value assigned to current output 1 is automatically assigned to PV.
SV, TV, QV
Freely selectable according to the setting options.