Setup
3.2 Process values
FST030
Function Manual, 04/2017, A5E35693328-AA
63
Related
Path
Setup → Process values → Density → Custom conversion factor
3.2.7.4
Decimal places
Description
Sets the decimal places for the process value displayed in the local display.
Setting Auto; No decimal place; 1 decimal place; 2 decimal places; 3 decimal places; 4 decimal
places; 5 decimal places; 6 decimal places
Default
2 decimal places
Related
Path
Setup → Process values → Density → Decimal places
3.2.7.5
Alarm and warning limits
3.2.7.5.1 Upper alarm limit
Description
If density value exceeds this limit an alarm is triggered.
Setting
0.00 kg/m³ to +10000.00 kg/m³
Default
900.00 kg/
Related
Path
Setup → Process values → Density → Alarm and warning limits → Upper alarm limit
3.2.7.5.2 Upper warning limit
Description
If density value exceeds this limit a warning is triggered.
Setting
0.00 kg/m³ to +10000.00 kg/m³
Default
890.00 kg/
Related
Path
Setup → Process values → Density → Alarm and warning limits → Upper warning limit