2-84
1010ENFM-3B
Section 2
Procedure to clear Active Memory only:
Turn off power (if it is currently on).
Press <F4> and keep it pressed while you turn on power.
The prompt: [Clr Active Memory? No] appears at the top of the screen
To access the [F4 Reset] option list press <Right Arrow>. Press <Down Arrow> to switch option
list to [Clr Active Memory? Yes]. To clear all Active Site Data (but not saved Site Setups). Press
<ENT>.
To restore operation, press <MENU> to access the installation menu. Create a new site setup or
recall a stored site setup.
Re-select any Meter Facilities items (e.g., RS-232 setup parameters).
Procedure to clear ALL Dynamic Memory:
Turn off power (if it is currently on).
Press <F4> and keep it pressed while you turn on power. The prompt: [Clr Active Memory?
No] appears at the top of the screen. Press <Down Arrow>. Note that the prompt switches to
[Clr Saved Data? No].
To access the F4 Reset option list press <Right Arrow>. Press <Down Arrow> to switch the option
list to [Clr Saved Data? Yes].
CAUTION: Before proceeding further it is essential to understand that this function eliminates
ALL data stored in RAM. This means that all saved site setups including the site data of
a flow-calibrated site will be erased! In addition, the entire Datalogger file plus any
custom factory or user-created pipe or transducer tables will be eliminated. The im-
pact of this is such that we strongly recommend that you consult technical service
before continuing with this procedure. Be aware that you will have to create a new
Site Setup, re-enter all site specific parameters including pipe or transducer tables,
plus all desired Meter Facilities entries.
To clear all Dynamic Memory press <ENT>.
Create a Site Setup before attempting to access other menu items.
To restore operation, press <MENU> to access the Installation Menu. Create a new site setup and
complete the installation procedure.
Re-select desired Meter Facilities items (e.g., RS-232 setup parameters).