CAUTION
In applications with strong vibrations, Siemens recommends remote mounting of the
transmitter!
4.3 Determining a location
Note
The sensor must always be completely lled with liquid.
Figure4-2 Correct installation with lled pipes
• Avoid the following installations
– Installation at the highest point in the pipe system
– Installation in vertical pipes with free outlet
Figure4-3 Wrong installation at high point
Figure4-4 Correct installation at low point before outlet
Installing/Mounting
4.3Determining a location
SITRANS FMS500