4.4 Inlet and outlet conditions
Inlet and outlet conditions
To achieve accurate 󹪞ow measurement it is essential to have straight lengths of inlet and outlet
pipes and a certain distance to pumps and valves.
It is also important to centre the 󹪞owmeter in relation to pipe 󹪞anges and gaskets.
Figure4-5 Inlet and outlet conditions
Installation in partially 󹪝lled pipes
For partially 󹪝lled pipes or pipes with downward 󹪞ow and free outlet the 󹪞owmeter should be
located in a U-tube.
Installation in large pipes
The 󹪞owmeter can be installed between two reducers (for example DIN 28545). At α ≤ 8° the
following pressure drop curves apply. The curves are applicable to water.
Installing/Mounting
4.4Inlet and outlet conditions
SITRANS FMS500
Operating Instructions, 01/2024, A5E52677520-AA 19