Before commissioning it must be checked that:
The device has been installed and connected in accordance with the guidelines provided
previous in this chapter and in Installing/Mounting (Page17).
5.5 Potting
If sensor is buried or permanently submerged, terminal box must be encapsulated with silicon
dielectric gel (non-toxic, transparent and self-healing gel).
NOTICE
Electrical connections
Do not pot meter before electrical connections have been made.
Mix the two components of the potting kit well and pour into terminal box.
Let cure for approximately 24 hours at approximately 25°C (77°F). Curing time increases by
100% per -10°C (-18°F).
Horizontal orientation Vertical orientation
Note
Gel can be penetrated with test instruments or be removed in case of cable replacement.
Connecting
5.5Potting
SITRANS FMS500
Operating Instructions, 01/2024, A5E52677520-AA 37