EasyManua.ls Logo

Siemens SITRANS FS230 - Page 16

Siemens SITRANS FS230
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Table 2-5 FSS200 Very High Temperature Sensors; Entity Parameters: Ui = 17.44 V, Ii = 461 mA, Pi = 1.00 W
MLFB Number Ci (nF) Li (uH) Temperature code
7ME3950aLAb 0 0 T5 for Ta ≤ 85°C
T4 for Ta ≤ 100°C
Where,
a = Approval: 5 (FM/CSA/ATEX/IECEX)
b = Size: 13,23,33,43,63,73,83
Note
Control drawing
* See Control drawing A5E37305975A
Installation variations
Note
Requirements for safe installation
The sensors can be installed in Zone 0, Div. 1 as Intrinsically Safe.
Standard remote installation with FST030 because the connection is certi󹪝ed Intrinsically
Safe.
Safety notes
2.3Use in hazardous areas
SITRANS FS130 & FS230
16 Compact Operating Instructions, 12/2023, A5E38755607-AF