4. Push cable through gland opening.
Figure4-15 Connection - SSL cable inside
5. Connect the four wires according to list below.
Terminal number Description Wire color (SSL cable)
17 Vcc+ Orange
18 Vcc- Yellow
19 Rx White
20 Tx Blue
6. Assemble and tighten cable gland.
Connecting sensor DSL
1. Remove lock screw and remove DSL lid.
2. Undo the 󹪞exible strap.
3. Disconnect sensor connection from DSL cassette.
4. Loosen mounting screw using a TX10 Torx driver and remove DSL cassette from housing.
5. Remove cap and ferrule from cable gland and slide onto cable.
6. Push cable through open gland; anchor cable screen and wires with clamp bar.
7. Remove terminal block from DSL cassette.
Connecting
4.7Connecting wall mount enclosure
SITRANS FS130 & FS230
Compact Operating Instructions, 12/2023, A5E38755607-AF 71