Additional Installation Instructions
B.3 Magnetic mounting
FSS200 clamp-on sensors
Installation Manual, 08/2017, A5E36255466-AC
111
3. Align the mounting pins of the clamp plate assembly with the frame and insert into the
frame.
4. Select the desired sensor hole and screw in the Clamping Screw.
Figure B-23 Clamp plate assembly
Mounting pins
sensor selection positioning holes
Locking nut
Clamping screw
5. Tighten the sensor clamping screw to hold the sensor firmly in place. Tighten the
clamping screw locking nut but not so tight that the magnetic frame lifts off the pipe.
Repeat procedure for the other sensor.
6. Fill connector end with Super Lube prior to connecting.
7. Apply Super Lube to the internal threads at the large end of the thread connector.
8. Observing the upstream and downstream orientation, attach the upstream (1A) and
Downstream (1B) cables to the sensors and make snug. Attach other ends to the
transmitter at 1A and 1B terminals of the transmitter.
9. Proceed to Sensor wiring (Page 48).
Mounting temperature sensors (Page 56)