Additional Installation Instructions
B.4 Hi Precision mount
FSS200 clamp-on sensors
Installation Manual, 08/2017, A5E36255466-AC
115
Note
Applying sealant - Refer to in figure above.
In environmentally challenging conditions it is an accepted practice to apply a heavy bead of
sealant material such as silicone caulking or other environment appropriate material between
the enclosure and the pipe surface that leaves an area at the lowest elevation free for
drainage.
Note
Always mount sensors above the horizontal plane of the pipe. Avoid mounting directly on top
or bottom of horizontal pipes.
Single enclosure Reflect mount
Sensor locations
Downstream sensor
Reference sensor (REF) or fixed sensor
Figure B-26 Single enclosure - Reference sensor location
Mounting straps for Reflect mount - Single enclosure
1.
Use the Wizard setup procedure to program the transmitter for the sensors that were
selected.
2. Determine the correct sensor spacing as indicated by the transmitter.