ID/
Icons
Diagnostic Action Comment
271 Specific gravity Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Lower alarm limit"
Only hydrocarbon
272 Standard specific gravity Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Upper alarm limit"
Only hydrocarbon
273 Standard specific gravity Value above warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter "Upper warning limit"
Only hydrocarbon
274 Standard specific gravity Value below warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter "Lower warning limit"
Only hydrocarbon
275 Standard specific gravity Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Lower alarm limit"
Only hydrocarbon
276 LiquIdent Value above alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Upper alarm limit"
Only hydrocarbon
277 LiquIdent Value above warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter "Upper warning limit"
Only hydrocarbon
278 LiquIdent Value below warning limit. Check process con‐
ditions or align limit to normal operation. Adjust
parameter "Lower warning limit"
Only hydrocarbon
279 LiquIdent Value below alarm limit. Check process condi‐
tions or align limit to normal operation. Adjust
parameter "Lower alarm limit"
Only hydrocarbon
Diagnostics and troubleshooting
9.8 Fault codes and corrective actions
SITRANS FST030 (Modbus)
120 Operating Instructions, 08/2018, A5E37367739-AD