EasyManua.ls Logo

Siemens SITRANS FST030 - Page 202

Siemens SITRANS FST030
356 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Modbus
register
Data type /
Size in bytes
Parameter Description Default value
[units]
(units register)
Value range /
Setting options
Access
level
8806 Unsigned / 2 Current mode Mode of current output. 0
0: 4-20 mA (3.5)
3.8-20.5 (22.6)
NAMUR
1: 4-20 mA (3.75)
4.0-20.8 (22.6) US
2: 4-20 mA (2.0)
4.0-20.5 (22.0)
3: 4-20 mA (2.0)
4.0-24.0 (25.0)
4: 0-20 mA (0.0)
0.0-20.5 (22.0)
5: 0-20 mA (0.0)
0.0-24.0 (25.0)
Read /
write
8866 Float / 4 Upper range
value volume
flow
Upper volume flow val‐
ue that is mapped to
the upper nominal
range boundary selec‐
ted by Current mode
(8806). Applicable if
Process value (8802) =
1.
0.004 • max.
volume flow
[m
3
/s]
(7500)
- Read /
write
8868 Float / 4 Lower range
value volume
flow
Lower volume flow val‐
ue that is mapped to
the lower nominal
range boundary selec‐
ted by Current mode
(8806). Applicable if
Process value (8802) =
1.
0
[m
3
/s]
(7500)
- Read /
write
8810 Float / 4 Upper range
value mass flow
Upper mass flow value
that is mapped to the
upper nominal range
boundary selected by
Current mode (8806).
Applicable if Process
value (8802) = 0.
0.2 • max.
mass flow
[kg/s]
(7400)
- Read /
write
8812 Float / 4 Lower range
value mass flow
Lower mass flow value
that is mapped to the
lower nominal range
boundary selected by
Current mode (8806).
Applicable if Process
value (8802) = 0.
0.0
[kg/s]
(7400)
- Read /
write
Modbus communication
A.9 Modbus holding registers tables
SITRANS FST030 (Modbus)
202 Operating Instructions, 08/2018, A5E37367739-AD

Table of Contents

Related product manuals