View 4 settings
Table A-76 View 4 settings
Modbus
register
Data type /
Size in bytes
Parameter Description Default value
[unit]
(unit register)
Value range /
Setting options
Access
level
9015 Unsigned / 2 Enable or disa‐
ble
Enable or disable view. 1
● 0: Disabled.
● 1: Enabled.
Read /
write
9010 Unsigned / 2 View Type of appearance. 8
● 1: Single value.
● 2: Active diagnostic
events.
● 3: Three values.
● 4: Totalizer.
● 6: 1 value and bar
graph.
● 7: 1 value and graph.
● 8: Six values.
● 9: Six diagnostic
values.
Read /
write
9027 Unsigned / 2 1
st
value Selection of the first val‐
ue.
1 See Selectable values
dependent on the view
type (Page 279).
Read /
write
9028 Unsigned / 2 2
nd
value Selection of the second
value. Depending on
the number of values.
2 See Selectable values
dependent on the view
type (Page 279).
Read /
write
9029 Unsigned / 2 3
rd
value Selection of the third
value. Depending on
the number of values.
3 See Selectable values
dependent on the view
type (Page 279).
Read /
write
9084 Unsigned / 2 4
th
value Selection of the fourth
value. Depending on
the number of values.
0 See Selectable values
dependent on the view
type (Page 279).
Read /
write
9085 Unsigned / 2 5
th
value Selection of the fifth val‐
ue. Depending on the
number of values.
11 See Selectable values
dependent on the view
type (Page 279).
Read /
write
9086 Unsigned / 2 6
th
value Selection of the sixth
value. Depending on
the number of values.
12 See Selectable values
dependent on the view
type (Page 279)..
Read /
write
8068 Unsigned / 2 Trend scale
mode
Scaling mode for the
graph and totalizer
view.
0
● 0: Automatic scaling.
● 1: Fixed scaling
defined by Trend
scale lower limit
(8070) and Trend
scale upper limit
(8072).
Read /
write
Modbus communication
A.9 Modbus holding registers tables
SITRANS FST030 (Modbus)
274 Operating Instructions, 08/2018, A5E37367739-AD