5. Connect ground to terminal and power to terminals L/+ and N/- in the manner shown
below at right using a screwdriver.
Note
The terminal box is detachable
For easier access unplug the terminal box. After connecting the wires, plug the terminal box
back in.
① L/
+
② N/
-
③ Protective Earth (PE)
AC connection DC connection
Power: 85 to 264 V AC, 47 to 63 Hz
Power: 19.2 to 28.8 V DC
6. Tighten cable gland.
7. Mount power supply protection cover and fasten protection cover screw.
5.6 Finishing the transmitter connection
Connection check-up
1. Check individual wire installation by tugging firmly.
2. Firmly tighten cable glands and insert blanking plugs in unused cable entries.
Connecting
5.6 Finishing the transmitter connection
SITRANS FST030 (Modbus)
48 Operating Instructions, 08/2018, A5E37367739-AD