Troubleshooting Symptom: (Zero flow and zero sound velocity)
Symptoms / Alarms Diagnostics Cause Remedy
Display shows no flow and
measured sound velocity is 0
m/s.
No Alarm codes
Path 1 state: 0-0
Path 2 state: 0-0
No paths installed
At least 1 path must be
installed for flow
measurement to start.
Follow instructions in
sensor setup wizard while
ensuring that at least 1
path is selected from the
installed paths multi-
selection menu.
Display shows no flow and
measured sound velocity is 0
m/s.
With Alarm codes:
14: Path1 No Signal
and / or
15: Path2 No Signal
and
34: Invalid flow meas.
RxGain > 60
No Rx signal present on
any path
Pipe not completely filled with
liquid.
Make sure that the pipe is
completely filled with
liquid, or relocate sensors
to a section of pipe that is
expected to be normally
full (see A2 (Page 95)).
Avoid installation of
sensors on top or bottom
of horizontal pipe.
Cable broken, detached
or attached to the wrong
F-connector channels.
Broken or misswired
cables.
Ensure that the sensor
cables are properly
connected and routed
between the sensors and
transmitter.
Verify continuity by
evaluating the DC
resistance between the
cable center conductor
and shield with the
sensors connected.
Incorrectly installed sensors
or incorrect setup parameter‐
ization.
Verify pipe dimensions
Verify sensor selection,
mounting configuration
and spacing.
Ensure that ultrasonic
couplant is applied
(grease or dry couplant)
(see A4 (Page 99)).
Damaged / defective sensors
or frontend hardware.
Verify operation of sensor
and frontend hardware using
a pipe simulator "Test block"
(see A1 (Page 93))
Diagnostics and troubleshooting
9.2 Troubleshooting Guide
SITRANS FST030 (Modbus)
Operating Instructions, 08/2018, A5E37367739-AD 89