Symptoms / Alarms Diagnostics Cause Remedy
Display shows no flow and
measured sound velocity is 0
m/s.
With Alarm code:
36: Configuration 2
No Rx signal present on
any path
The pipe dimensions are too
small for the selected sensor
size or frequency.
Verify pipe dimensions,
fluid selection, and
sensor selection.
Application may not be
suitable for the purchased
sensor size.
Display shows no flow and
measured sound velocity is 0
m/s.
With Alarm code:
34: Invalid flow meas.
Rx signal is intermittently
present but does not show a
proper leading baseline:
For comparison Rx signal
with proper leading baseline:
Incorrectly installed sensors
or incorrect setup parameter‐
ization.
Verify actual pipe
dimensions against
entered values.
Verify sensor selection,
mounting configuration
and spacing (see A4
(Page 99)).
Incorrect Expected fluid
sound velocity.
A very large error in the
expected fluid sound
velocity (or sound speed)
can result in a large
sensor spacing error and
cause the meter to search
for the signal in the wrong
transit-time range.
Verify that the entered
expected sound velocity
is correct for the
application.
Pipe is mostly empty causing
meter to operate on pipe
noise.
Make sure that the pipe is
completely filled with
liquid, or relocate sensors
to a section of pipe that is
expected to be normally
full.
Avoid installation of
sensors on top or bottom
of horizontal pipe (see A4
(Page 99)).