Parameter lists
A.4 Meter setup
SITRANS FUS080 transmitter
Operating Instructions, 07/2012, A5E03059912-02
79
Modbus
Address
Parameter Default
value
Value
range
Access
level
Description
Pipe data
40181 300 Number of paths Product
dependent
2 R/W Number of paths on the sensor
40149 302 Max sample frequency 15 15 Read The maximum frequency for flow
measurement
40147 303 Sample frequency 0.5 15 HW key The frequency at which the flow is
measured
40167 304 Rn for path 1 Product
dependent
3.4E+38
m
Read R-factor for path 1.
For FUS080/SONOKIT (retrofit)
installations this parameter will
automatically be calculated by the pipe
geometric tool (see 'Pipe geo.
assistant' in 'Device' menu).
40169 305 Rn for path 2 Product
dependent
3.4E+38
m
Read R-factor for path 2.
For FUS080/SONOKIT (retrofit)
installations this parameter will
automatically be calculated by the pipe
geometric tool (see 'Pipe geo.
assistant' in 'Device' menu).
Valid only for 2-path solutions
40171 306 Inner pipe diameter Product
dependent
0.05 ...
1.200 m
Read Inner pipe diameter in meters.
For FUS080/SONOKIT (retrofit)
installations this parameter will
automatically be calculated by the pipe
geometric tool (see 'Pipe geo.
assistant' in 'Device' menu).