3 INSTALLATION
30
SITRANS FX300
www.siemens.com/flow 06/2009 - SFIDK.PS.050.F2.02 A5E02100423
3.6.3 Installing devices in sandwich design
Push the first bolt 3 through the hole 5 of both flanges.
Screw on the nuts and washers to both ends of the bolt 3 but do not tighten them.
Install the second bolt through the holes 4.
Place the measuring sensor 1 between the two flanges.
Insert the gaskets 6 between measuring sensor 1 and flanges and align them.
Check that the flange is concentric.
Install the remaining bolts, washers and nuts. Do not yet tighten the nuts.
Turn the centring ring 2 in a counter-clockwise direction and align the device.
Check that the gaskets 6 are concentric; they must not protrude into the pipe cross-section.
Now tighten all nuts bit by bit alternately across the diagonal.
Figure 3-13: Installation using centering ring
1 Measuring sensor
2 Centering ring
3 Bolts with fixing nuts
4 Hole
5 Hole
6 Gasket