OPERATION 6
97
SITRANS FX330
www.siemens.com/flow08/2022 - A5E36751269-AG EN
M Electronics Information
Faulty Converter Oscillator The converter oscillator is
working out of range.
Perform a hard reset. If the
message recurs, contact the
manufacturer.
Faulty conv. voltage ref. Reference voltage in
converter module out of
specification.
C Process
Heater / Cooler Error During net heat metering
the difference in
temperature is reversed.
Check net heat metering
setup.
Check temperature input
configuration.
Table 6-40: Status messages
Status
type
Event group Single event Description Actions to eliminate the
event