EasyManua.ls Logo

Siemens SITRANS I100 7NG4124-0AA00 - Engineering; Max. Ambient Temperatures; Power Dissipation

Siemens SITRANS I100 7NG4124-0AA00
24 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
english
SITRANS I100
13
Operating instruction A5E02591611-02
5 Engineering
5.1 Max. ambient temperatures
The SITRANS I100 isolators can be used over a wide temperature range. Depending on
the isolator version and installation method different maximum ambient temperatures may
result.
Ventilation:
without ventilation
Installation:
Single unit
DIN
-
rail
Orientation:
any
vertical
horizontal
Type:
SITRANS I100
70 °C
60 °C 65 °C
Ventilation:
with ventilation
Installation:
Single unit
DIN
-
rail
Orientation:
any
vertical
horizontal
Type:
SITRANS I100
70 °C
70 °C 70 °C
5.2 Power dissipation
Data sheets are describing the maximum power dissipation in standard operation (output
20 mA; power supply 24 V DC; load = 250
). In practice not all isolators are working with
full load. Therefore engineering is done typically with an average power dissipation of 70 %
(P
70%
).
Type max. power dissipation 70 % power dissipation
SITRANS I100 1.5 W 1.1 W