7ML19985JF02 SITRANS LR250 (PROFIBUS PA) – INSTRUCTION MANUAL Page 85
mmmmm
Parameters
3.5.8. Acknowledge Status
Acknowledges the status of the Maintenance limits.
In SIMATIC PDM, open the menu View – Device Status, click on the
Maintenance tab and check the Sensor Lifetime Status window.
3.5.9. Acknowledge
Acknowledges the current limit status.
To acknowledge an alert via SIMATIC PDM:
a) Open the menu View – Device Status and click on the Maintenance tab.
b) In the Sensor Lifetime section click on Acknowledge Warnings.
To acknowledge an alert via the handheld programmer:
a) Press RIGHT arrow twice to open parameter view and activate Edit
Mode.
b) Press RIGHT arrow to acknowledge the alert.
3.6. Condensed Status Setup
When Condensed Status Mode (3.6.1.) is enabled, you can select the level of severity
of errors, and tailor a device response appropriate for your particular process.
•In Event Index (3.7.1.) you can select a particular event or error by means of its
index number.
•In Event Status (3.7.2.) you can assign a status to the selected event.
•In Event Diagnosis (3.7.3.) you can assign a diagnosis to the selected event.
3.6.1. Condensed Status Mode
Select Yes or No to enable/disable Condensed Mode.
3.6.2. Features supported (View only)
Features supported are:
• Condensed Diagnostics
• Extended Diagnostics
• Application Relationships
3.6.3. Features enabled (View only)
Lists those features that have been enabled.
Note: When cyclic communication is in progress, Condensed Status Mode
cannot be changed.
Options
No (disabled)
*
Yes (enabled)