10.3.1 Enclosure open
WARNING
Enclosure open
Risk of e
xplosion in hazardous areas.
To open the device in a hazardous area, isolate the device from power.
Exception: Devices exclusively having Intrinsic safety (Exi) may be opened in an energized
state in hazardous areas.
WARNING
Improper connection after maintenance
Risk of e
xplosion in areas subject to explosion hazard.
Connect the device correctly after maintenance.
Close the device after maintenance work.
Refer to Connecting (Page49).
WARNING
Hot, toxic, or corrosive process media
Risk of injur
y during maintenance work.
When working on the process connection, hot, toxic, or corrosive process media could be
released.
As long as the device is under pressure, do not loosen process connections and do not
remove any parts that are pressurized.
Before opening or removing the device ensure that process media cannot be released.
CAUTION
Hot surfaces
Risk
of burns during maintenance work on parts having surface temperatures exceeding 70 °C
(158°F).
Take corresponding protective measures, for example, by wearing protective gloves.
After carrying out maintenance, remount touch protection measures.
Service and maintenance
10.3Maintenance and repair work
SITRANS LR500 series with mA/HART
Operating Instructions, 03/2024, A5E51099898-AA 151