Page 68 COMMUNICATIONS FOR SITRANS LUT400 (HART) – MANUAL A5E33701270
mmmmm
Operation - AMS
Changing parameter settings using AMS Device Manager
1) Adjust parameter values in the parameter value field in Configure/Setup view, then
click on Apply to write the new values to the device. The parameter field will display
in yellow until the value has been written to the device.
2) Click on OK only if you wish to update all parameters and exit AMS.
Configure/Setup View
Operation
Navigate to Configure/Setup > Operation for access to:
Select Analog Output:
• Select analog output and current output function
Auto Zero Head
• View current head measurement, set actual head value and set zero head
offset
Auto Sensor Offset
• View current distance measurement, set desired distance for auto sensor
offset, and set sensor offset
Auto Sound Velocity
• View current distance measurement, set desired distance for auto sound
velocity, and set sound velocity at 20 degrees Celsius.
Real Time Clock
• Set current date and time
Tests
• Perform loop-test
Note: For a complete list of parameters, see
AMS Menu Structure
on page 78.