SITRANS LUT400
Operating Instructions, 07/2019, A5E33329501-AC
169
9.2.1.3 Sensor Mode (2.1.3.)
Menu number
2.1.3.
visible on LUT430 (Pump and Flow model), and LUT440 (OCM model).
Sets the type of measurement required for the application.
Options (Mode)
Description
Reference point
* LEVEL Distance to material surface Low Calibration Point (process
empty level)
SPACE High Calibration Point (process
full level)
DISTANCE Sensor Reference Point
VOLUME Volume of material in volumetric units
(based on level)
Low Calibration Point
HEAD
1
Distance to material surface
Zero Head
FLOW
1
Flowrate in an open channel in Flowrate
Units
Zero Head (zero flow level)
1)
Option available only on LUT430, LUT440.
Distance
Sensor reference point
High calibration point
Space
Level
Low calibration point
Far range