SITRANS LUT400
Operating Instructions, 07/2019, A5E33329501-AC
233
Zero Head Offset (2.15.3.5.)
The difference (positive or negative) between Low Calibration Point and zero head (level at
zero flow), defined in Units (2.1.1.) (Page 168).
Values
Range: -60.000 to 60.000
Default: 0.000
This feature can be used for most weirs and some flumes (e.g. Palmer Bowlus) where the
zero reference is at a higher elevation than the channel bottom.
Low calibration point
Zero head
Transducer
Material level at zero flow
PMD base
Zero head offset
Flowrate Decimal (2.15.3.6.)
The maximum number of decimal places to be displayed.
Options
*
NO DIGITS
no digits after the decimal point
1 DIGIT
1 digit after the decimal point
2 DIGITS
2 digits after the decimal point
3 DIGITS
3 digits after the decimal point
Flowrate Units (2.15.3.7.)
The volume units used to display total flow.
Options
*
L/S (Litres per second)
L/MIN (Litres per minute)
CUFT/S (Cubic feet per second)
CUFT/D (Cubic feet per day)
GAL/MIN (US Gallons per minute)
GAL/D (US Gallons per day)