Zero Head Offset (2.15.3.5.)
The difference (positive or negative) between Low Calibration Point and zero head (level at
zero flow), defined in Units (2.1.1.) (Page 168).
This feature can be used for most weirs and some flumes (e.g. Palmer Bowlus) where the
zero reference is at a higher elevation than the channel bottom.
Material level at zero flow
Flowrate Decimal (2.15.3.6.)
The maximum number of decimal places to be displayed.
no digits after the decimal point
1 digit after the decimal point
2 digits after the decimal point
3 digits after the decimal point
Flowrate Units (2.15.3.7.)
The volume units used to display total flow.
L/MIN (Litres per minute)
CUFT/S (Cubic feet per second)
CUFT/D (Cubic feet per day)
GAL/MIN (US Gallons per minute)
GAL/D (US Gallons per day)