Error
No.
Error text
Remedy text
Comment Output
status
Input
status
42
P42 - Current output
Check cables Current loop is disconnected or the loop resistance is
too big
Active Active
43
P43 - Internal error
Switch off and on Too many errors occurred at the same time.
Some errors are not detected correctly
Active Active
44
P44 - CT SENSORPROM®
Replace SENSORPROM
®
unit has been used as CT version Active Active
49
P49 - Protection violation
Switch off and on
Internal protection of the device has been violated. Active
60
F60 - CAN comm. error
Transmitter/AOM CAN bus communication error. An add-on module,
the display module or the transmitter is defective
Zero Inactive
61
F61 - SENSORPROM® error
Replace It is not possible to rely on the data in SENSOR‐
PROM
®
unit anymore
Active Active
62
F62 - SENSORPROM® ID
Replace The SENSORPROM
®
unit ID does not comply with
the product ID.
The SENSORPROM
®
unit is from another type of
product SITRANS F C, SITRANS F US etc.
Zero Inactive
63
F63 - SENSORPROM®
Replace It is not possible to read from the SENSORPROM
®
unit anymore
Active Active
70
F70 - Coil current
Check cables Coil excitation has failed Active Active
71
F71 - Internal error
Replace transmitter Internal conversion error in ASIC Active Active
Alarm, error, and system messages
8.2 List of error numbers
56
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018, A5E02083319-AH