Technical data
11
11.1 Technical specifications
Technical specifications for MAG 6000 I standard and Ex version
Mode of operation and design Measuring principle Electromagnetic with pulsed constant field
Empty pipe Detection of empty pipe (special cable re‐
quired in remote mounted installation)
Excitation frequency Depends on sensor size
Electrode input impedance > 1 x 10
14
Ω
Input Digital input 11 to 30 V DC, R
i
= 4.4 k
Ω
Activation time 50 ms
Current I
11 V DC
= 2.5 mA, I
30 V DC
= 7 mA
Output Current output
Signal range 0 to 20 mA (active/passive)
Active current output R
max
500 Ω @ V
max
30 V DC
Passive current output R
min
0 Ω @ V
min
12 V DC
R
max
500 Ω @ V
max
30 V DC
Time constant 0.1 to 30 s, adjustable
Digital output
Frequency 0 to 10 kHz, 50% duty cycle (uni/bidirectional)
Time constant 0.1 to 30 s, adjustable
Pulse (passive) 3 to 30 V DC, max. 110 mA (Ex version: max.
30 mA), 200 Ω ≤ R
i
≤10 kΩ (powered from
connected equipment)
Time constant 0.1 to 30 s, adjustable
Relay output
Time constant Changeover relay, same as current output
Load 30 VAC/2 A, 24 VDC/1 A
Low flow cut off 0 ... 9.9% of maximum flow
Galvanic isolation All inputs and outputs are galvanically isolated PELV circuits with 60 VDC isolation from
each other and ground. Exceptions are sensor connections that are connected to
ground.
Max. measuring error
(incl. sensor and zero point)
MAG 6000 I/MAG 6000 I Ex de ±0.2% ± 1 mm/s
1)
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018, A5E02083319-AH
65