Technical specifications for MAG 6000 I standard and Ex version
Rated operation conditions Ambient temperature
Operation Standard version
-25 to +60 °C (-13 to +140 °F)
Ex version:
-25 to +60 °C (-13 to +140 °F)
Storage -40 to +70 °C (-40 to +158 °F)
Mechanical load 18...1000 Hz random in x, y z, directions for 2
hours according to EN 60068-2-36 Transmit‐
ter: 1.14 grms
Protection degree IP67/NEMA 4X to IEC 60529 and DIN 40050
(1 mH
2
O 30 min.)
EMC performance EN 61326
Display and keypad Totalizer Two eight-digit counters for forward, net or re‐
verse flow
Display Background illumination with alphanumeric
text, 3 x 20 characters to indicate flow rate,
totalized values, settings and faults;
Reverse flow indicated by negative sign
Keypad Capacitive touch keypad with LED light for
feedback indication
Time constant Time constant as current output time constant
Design Enclosure material Die cast aluminum with corrosion-resistant Ba‐
sic Polyester power coating (min 60 µM)
Wall mounting bracket Bracket enclosed for remote mounting
Dimensions See dimensional drawings
Weight See dimensional drawings
Power supply Standard version: 100 to 230 VAC, 47 to 63 Hz
20 to 27 VDC
Ex version: 100 to 230 VAC, 47 to 63 Hz
20 to 27 VDC
Power consumption 230 V AC: 21.5 VA
24 V DC: 12 W, I
N
= 380 mA, I
ST
= 1.5 A (10 ms)
Certificates and approvals Standard version CE
C-tick
FM Class I, Div 2
FM Class I, Zone 2
CSA Class I, Div 2
CSA Class I, Zone 2
Ex version Ex d e [ia] ia IIC T6 Gb Ex tD A21 IP67 T85°C
Ta -25°C to +60°C
FM Class I, II, III, Div 1
3)
FM Class I, Zone 1/21
CSA Class I, Zone 1/21
Technical data
11.1 Technical specifications