11.3 Accuracy
For accuracy reference conditions, see below.
)ORZPHWHUXQFHUWDLQW\
V: Flow velocity
E: Meter uncertainty as a percentage of measured value
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.2 ± 1 mm/s *
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.125/v) % of measured value *
* MAG 5100 W (Order No. 7ME652...) with DN 350 ... 1200
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.2 ± 2.5 mm/s
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.25/v) % of measured value
10
2 3 4 5 6 7 8 9 10 11 12 [m/s]
[ft/s]
[±% E]
0,0
0,2
0,4
0,6
0,8
1,0
1,2
1,4
3.30 6.6 9.9 13 16 20 23 26 30 33 36 39
Figure 11-2 MAG 6000 I / MAG 6000 I Ex de with MAG 1100 (not PFA), MAG 1100 HT, MAG 1100
F (not PFA), MAG 5100 W, MAG 3100 P, MAG 3100 and MAG 3100 HT
Reference conditions
(ISO 9104 and DIN/EN 29104)
A calibration certificate is shipped with every sensor and calibration data are stored in
SENSORPROM
®
memory unit.
Medium temperature 20 °C ± 10 K (68°F ± 18 °F)
Ambient temperature 25 °C ± 10 K (68°F ± 18 °F)
Supply voltage U
n
± 1%
Warming-up time 30 minutes
Incorporation in conductive pipe section
Inlet section
Outlet section
10 x DN (DN ≤ 1200/48")
5 x DN (DN > 1200/48")
5 x DN (DN ≤ 1200/48")
3 x DN (DN > 1200/48")
Flow conditions Developed flow profile
Reference conditions for sensor calibration
Current output As pulse output ± (0.1% of actual flow + 0.05% FSO)
Effect of ambient temperature
Display/frequency/pulse output
Current output
< ± 0.003% / °C act.
< ± 0.005% / °C act.
Technical data
11.3 Accuracy
68
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018, A5E02083319-AH