EasyManua.ls Logo

Siemens sitrans PROBE LU - Cyclic vs. Acyclic Data Transfer

Siemens sitrans PROBE LU
117 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 18 SITRANS Probe LU (HART) – OPERATING INSTRUCTIONS A5E32337695
Operation
Handheld Programmer: function keys in RUN mode
Certain functions can be accessed directly from RUN mode by using specific keys.
PROGRAM Mode
Programming
• Set parameters to suit your specific application.
• Activate PROGRAM mode at any time, to change parameter values and set
operating conditions.
• For local programming, use the Siemens handheld programmer.
• For programming from a distance, use either a PC running SIMATIC PDM, or a HART
handheld communicator.
Display
Key Run Mode
mA output value displayed in auxiliary reading field
Internal enclosure temperature displayed in auxiliary reading field (P343).
Parameter for auxiliary readings
1
1.
Press
plus three-digit parameter number, sets parameter to show in the auxiliary display.
Displays the value representing Echo Confidence (P805).
Toggle between Units and % on reading display
Initiate and complete PROGRAM mode access
Measurement key displays distance in auxiliary reading field.
Note: See
Accessing a parameter,
on page 21, for detailed instructions.
1
2
3
4
1 – Primary Reading (displays parameter value)
2 – Secondary Reading (displays parameter
number)
3 – Programming indicator
4 – Auxiliary Reading (displays parameter names
for parameters P001 to P010, if a language is
selected. It displays the index value for indexed
parameters, such as P054.)

Table of Contents

Other manuals for Siemens sitrans PROBE LU

Related product manuals