Page 46 SITRANS Probe LU (HART) – OPERATING INSTRUCTIONS A5E32337695
Parameters
P214 4 mA Output Trim
Calibrates the 4 mA output.
P215 20 mA Output Trim
Calibrates the 20 mA output.
Installation Records (P300 to P346)
P300 Temperature, transducer maximum
Shows the highest temperature encountered (in degrees C), as measured by the
temperature sensor in the transducer (if applicable).
Press CLEAR then ENTER , to reset the log after a short circuit on the transducer
wiring.
P341 RUN Time
Displays the number of uninterrupted 24 hour periods that the device has been operating.
P341 is updated once a day.
• If power is cycled before 24 hours have passed, the run time will not be updated.
• If an instrument is powered down on a regular basis, P341 will not have an
accurate value.
Note: This parameter is for use only by Siemens service personnel.
Note: This parameter is for use only by Siemens service personnel.
Primary Index
Transducer
Values (view only)
Range –50 to 150 oC (view only)
Default –50 (oC)
Values (view only)
Range 0 to 99999 (days)
Default: 0
Related
P342: Power-on Resets