Operation
7.3 Menu system
SITRANS SL
Operating Instructions, 12/2010, A5E01132948-04
123
About SITRANS SL
Informative display informs about the following system parameters:
Analyzer serial number
Analyzer revision
Application name
Software Version
MLFB (order) number
Transmitter FPGA
Receiver FPGA
Transmitter PLD
Junction PLD
Calibrate
CAUTION
Never use this function without having contacted Siemens service first!
Inappropriate use of this function may seriously affect the accuracy of the analyzer.
After selecting one of the gas components to be calibrated the calibration menu is displayed
providing the following options:
Measured value - displays currently measured value
Setpoint – edit setpoint
Calibrate – perform (or cancel) calibration
Reset
Reset the analyzer
This action requires confirmation before it is started.
Analog output test
Select output to test:
Analog out 0
Analog out 1
Once an output has been selected a confirmation is requested before test mode is entered.
After confirmation the output current is displayed and can be edited.