syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Accessories System Description
B.7-10
ECG electrodes
B.7
The ECG electrodes are connected to the receptacle at the
patient table. (→ Page B.3-2 Operating elements and connec-
tors)
B.7
When applying the ECG electrodes to the patient, the color cod-
ing according to IEC or AHA is recommended.
B.7
Placement for Europe ver-
sion
B.7
B.7
Color coding according to IEC
B.7
Red electrode - on the right mid-clavicular line, directly
below the clavicle
Yellow electrode - on the left mid-clavicular line, 6 or 7
intercostal space
Green electrode - right mid-clavicular line, 6 or 7 intercos-
tal space
Placement for USA version
B.7
B.7
Color coding according to AHA
B.7
White electrode - on the right mid-clavicular line, directly
below the clavicle
Black electrode - on the left mid-clavicular line, 6 or 7 inter-
costal space
Red electrode - right mid-clavicular line, 6 or 7 intercostal
space