System Description Accessories
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
B.7-11
0.0
Cleaning
B.7
The ECG cable should be cleaned with isopropyl alcohol or
another mild disinfectant only. Other agents, especially sub-
stances which contain acetone should not be used. This acces-
sory may not be steam sterilized.
B.7
Respiratory Gating System (optional)
B.7
For motion artifact-free CT images in the chest region, a respi-
ratory triggering and gating function is available.
B.7
The respiratory triggering and gating function requires the use
of the Respiratory Gating System (AZ-733V, Anzai Medical: res-
piratory belt with pressure sensor, signal converter, and sensor
unit). It is not possible to use both triggering and gating simul-
taneously.
B.7
i
.7
B
Please contact your local Siemens representative to find out
whether the Respiratory Gating system is released for your CT
system.
B.7
B
.7
B
Improper positioning of the respiratory belt can cause image
artifacts. Position the belt outside the scan region.
B.7