System Handling Operation of the Patient Table
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
C.2-9
0.0
Retracting the patient table
manually
C.2
In an emergency or power failure, you can position the table top
manually.
C.2
Moving the table top horizon-
tally
C.2
You can use the handle at the end of the table to move the table
top out of the gantry.
C.2
C.2
Press the lever towards the end of the table top and pull the
table top out of the gantry.
B
.2
C
Before you continue work on the system, you must lock the tab-
letop back into its original position.
C.2