0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Switching the System On and Off Operation
D.1-20
Gantry temperature outside permissible
range
D. 1
Accurate measurements are only possible if the temperature
inside the gantry is within a certain range.
D. 1
B
D.1
If the temperature in the gantry falls below a specific limit, the
quality of the images is no longer guaranteed because calibra-
tion is no longer correct.
D. 1
Overheating or low tempera-
ture
D. 1
If the temperature of the gantry exceeds the permissible range,
countdown starts. A temperature warning is displayed.
D. 1
Recovery
D. 1
If the temperature moves back into the permissible range dur-
ing countdown, the countdown stops. You can then return to
the user interface.
D. 1
Shutdown
D. 1
When the countdown has finished, measurement is disabled. A
message is displayed in the status line. Measurement remains
switched off as long as the temperature is outside the permissi-
ble range.
D. 1