Quality Assurance CTDI Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.4-3
0.0
Positioning aid
F.4
To position the 16 cm CTDI head phantom use the CFK head
holder and, if necessary, the supplied thin pads to immobilize it.
The 32 cm CTDI body phantom has to be placed on the patient
table and fixed with the pads, if necessary.
F.4
F.4
Dosimeter
F.4
Measurement is performed with the dosimeter, a dose measur-
ing chamber with the necessary accessories, and a thermome-
ter.
F.4
i
.4
F
The dosimeter is not supplied with the system.
F.4
B
.4
F
Please follow the instructions for use supplied by the manufac-
turer of the measuring device.
F.4
Please note that the lead connecting the dosimeter and the ion-
ization chamber is very sensitive.
F.4