AbdThinSlice 2
nd
reconstr.
kV 120
Effective mAs 140
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 12.0 mm
Rotation time 0.5 sec.
Kernel B30f B20f
Increment 5.0 mm 0.7 mm
Image order cr-ca
CTDI
w
10.9 mGy
Effective dose Male: 3.4 mSv
Female: 4.1 mSv
If you want to use this protocol for a two phase study,
repeat the same protocol as mentioned below, and
chose start delay time for arterial phase as 20-25 s. In
this case, the thin slice reconstruction can also be used
for postprocessing. But do not administer oral contrast
medium, as this impairs the editing of MIP/SSD/VRT
images. Water could be used instead if necessary.
You can repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E. g. when both non-contrast and contrast studies are
required.
MPRthick: AbdThinSlice coronal
Image thickness 3
Distance between images 3
Number of images 50
The 2
nd
reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined MPR
range, linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay 50 65 sec.
Flow rate 4 ml/sec.
Total amount 100 ml
119
Abdomen