129
Pelvis
HipSpi 2
nd
reconstr.
kV 120
Effective mAs 150
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 12.0 mm
Rotation time 1.0 sec.
Kernel B30s B60s
Increment 5.0 mm 0.7 mm
Image order cr-ca
CTDI
w
11.7 mGy
Effective dose Male: 2.2 mSv
Female: 2.0 mSv
In case of 3D study only, mAs can be reduced up to
50%, images should be reconstructed with 50% over-
lapping and kernel B10.
If only one side is examined, it is advisable to enter the
side in the comment line.
MPRthick: HipSpi coronal
Image thickness 2
Distance between images 3
Number of images 30
The 2
nd
reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined range,
linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.