131
Pelvis
PelvisRoutine
kV 120
Effective mAs 140
Slice collimation 1.5 mm
Slice width 5.0 mm
Feed/Rotation 24.0 mm
Rotation time 0.5 sec.
Kernel B30f
Increment 5.0 mm
Image order cr-ca
CTDI
w
9.8 mGy
Effective dose Male: 3.3 mSv
Female: 4.8 mSv
* If the examination performed requires a full urinary
bladder, wait at least 3 minutes following IV admini-
stration of the contrast medium.
Contrast medium IV injection
Start delay 50 sec.*
Flow rate 2.0 – 3.0 ml/sec.
Total amount 100 – 120 ml