135
Spine
ObeseSeq
Indications:
Sequence mode for spine studies of the obese
patients when higher dose are needed, e. g. prolapse,
degenerative changes, trauma, tumors etc.
ObeseSeq
kV 120
Effective mAs 400
Slice collimation 0.75 mm
Slice width 1.5 mm
Feed/Scan 9.0 mm
Rotation time 1.5 sec.
Kernel B30s
Cycle time 3.0 sec.
Image order cr-ca
CTDI
w
33.6 mGy
Effective dose Male: 0.7 – 0.9 mSv
Female: 1.4 – 2.9 mSv