Spine
ThinSliceSpi
Indications:
Thin slice spiral mode for spine studies when high
resolution for fine structure is required.
ThinSliceSpi 2
nd
reconstr.
kV 120
Effective mAs 330
Slice collimation 0.75 mm
Slice width 3.0 mm 1.0 mm
Feed/Rotation 6.0 mm
Rotation time 0.75 sec.
Kernel B60s B60s
Increment 3.0 mm 0.6 mm
Image order cr-ca
CTDI
w
27.7 mGy
Effective dose Male: 6.6 mSv
Female: 13.2 mSv
MPRthick: ThinSliceSpi sagittal
Image thickness 3
Distance between images 3
Number of images 20
For the 2
nd
reconstruction the Autoload into the MPR
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MPR, and a coronal
MPRthick Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.