30
Children
HeadSpi
Indications:
Spiral mode for routine head studies, e. g. tumors,
hydrocephalus, hemorrhaging, abnormalities, etc.
HeadSpi
kV 120
Effective mAs *
Slice collimation 1.5 mm
Slice width 4.0 mm
Feed/Rotation 12.2 mm
Rotation time 0.75 sec.
Kernel C30s
Increment 4.0 mm
Image order ca-cr
Contrast medium IV injection
Start delay exam dependent
Flow rate dependent upon needle size/Access site
Total amount 1 – 2 ml per kg of body weight
* The mAs should be adjusted to the age of the child.
* The conversion factor for an 8-week-old,
and a scan range of 90 mm was used.
** The conversion factor for a 7-year-old, 260 mAs
and a scan range of 110 mm was used.
Body Weight kV mAs CTDI
w
Effective dose
(mGy) (mSv)
< 6 months 120 90 20.8 Male: 2.1*
Female: 2.1*
6 months – 120 150 34.8
3 years
3 years – 120 220 51.0 Male: 5.4**
6 years Female: 5.2**